Figure 5

image

Fig. 5. Results of the check of the equilibrator temperature probes from systems "A" through "G": Shown are the deviations of the measured temperatures from the reference temperature. Equilibrator probe and reference probe were kept together in the same water bath until readings became stable. Also shown are the linear correction lines that were applied to temperature readings of that particular system.