Carbon Dioxide, Hydrographic, and Chemical Data Obtained During the R/V Ronald H. Brown Repeat Hydrography Cruise in the Atlantic Ocean: CLIVAR CO2 Section A16S_2005 (11 January - 24 February, 2005)

NDP-087 (2006) ORNL/CDIAC-151

data Download the Data and ASCII Documentation files of NDP-087 PDF Download a PDF of NDP-087

image

Contributed by
R. Wanninkhof,1 S.C. Doney,2 E. Peltola,1 R.D. Castle,1 F.J. Millero,4 J.L. Bullister,3 D.A. Hansell,4 M.J. Warner,5 C. Langdon,4 G.C. Johnson,3 and C.W. Mordy3

Prepared by
Alexander Kozyr
Carbon Dioxide Information Analysis Center, Oak Ridge National Laboratory, Oak Ridge, Tennessee, U.S.A.

1Atlantic Oceanographic and Meteorological Laboratory, NOAA, Miami, FL
2Woods Hole Oceanographic Institution, Woods Hole, MA
3Pacific Marine Environmental Laboratory, NOAA, Seattle, WA
4Rosenstiel School of Marine and Atmospheric Science, University of Miami, Miami, FL
5University of Washington, Seattle, WA

Contents

Abbreviations and Acronyms

AOML  -  Atlantic Oceanographic and Meteorological Laboratory
BNL  -  Brookhaven National Laboratory
CDIAC  -  Carbon Dioxide Information Analysis Center
CFC  -  chlorofluorocarbon
CLIVAR  -  Climate Variability (Program)
CO2  -  carbon dioxide
CRM  -  certified reference material
CTD  -  conductivity, temperature, and depth sensor
DOC  -  dissolved organic carbon
DOE  -  U.S. Department of Energy
EXPOCODE  -  expedition code
fCO2  -  fugacity of CO2
GCOS  -  Global Climate Observing System
GCRP  -  Global Change Research Program
GOOS  -  Global Ocean Observing System
JGOFS  -  Joint Global Ocean Flux Study
LADCP  -  Lowered acoustic Doppler current profiler
LDEO  -  Lamont-Doherty Earth Observatory
MLR  -  multilinear regression
NDP  -  numeric data package
NDIR  -  nondispersive infrared analyzer
NOAA  -  National Oceanic and Atmospheric Administration
NSF  -  National Science Foundation
ORNL  -  Oak Ridge National Laboratory
pCO2  -  partial pressure of CO2
PMEL  -  Pacific Marine Environmental Laboratory
QA  -  quality assurance
QC  -  quality control
RSMAS  -  Rosenstiel School of Marine and Atmospheric Science
R/V  -  research vessel
SIO  -  Scripps Institution of Oceanography
SOMMA  -  single-operator multi-parameter metabolic analyzer
SST  -  sea surface temperature
TALK  -  total alkalinity
TOS  -  total organic carbon
TN  -  total nitrogen
TCO2  -  total inorganic carbon
UCI  -  University of California, Irvine
UW  -  University of Washington
WHP  -  WOCE Hydrographic Program
WOCE  -  World Ocean Circulation Experiment

Abstract

Wanninkhof, R., S.C. Doney, E. Peltola , R.D. Castle, F.J. Millero, J.L. Bullister, D.A. Hansell, M.J. Warner, C. Langdon, G.C. Johnson, and C.W. Mordy. 2006. Carbon Dioxide, Hydrographic, and Chemical Data Obtained During the R/V Ronald H. Brown Repeat Hydrography Cruise in the Atlantic Ocean: CLIVAR CO2 Section A16S_2005 (11 January - 24 February, 2005), ed. A. Kozyr. ORNL/CDIAC-151, NDP-087. Carbon Dioxide Information Analysis Center, Oak Ridge National Laboratory, U.S. Department of Energy, Oak Ridge, Tennessee, 38 pp. doi: 10.3334/CDIAC/otg.ndp087.

This report presents methods, and analytical and quality control procedures for salinity, oxygen, nutrient, inorganic carbon, organic carbon, chlorofluorocarbon (CFC), and bomb 14C system parameters performed during the A16S_2005 cruise, which took place from January 11 to February 24, 2005, aboard research vessel (R/V) Ronald H. Brown under the auspices of the National Oceanic and Atmospheric Administration (NOAA). The R/V Ronald H. Brown departed Punta Arenas, Chile, on January 11, 2005, and ended its cruise in Fortaleza, Brazil, on February 24, 2005. The research conducted was one of a series of repeat hydrography sections jointly funded by NOAA and the National Science Foundation as part of the CLIVAR/CO2/repeat hydrography/tracer program. Samples were taken from 36 depths at 121 stations.

The data presented in this report include the analyses of water samples for total inorganic carbon (TCO2), fugacity of CO2 (fCO2), total alkalinity (TALK), pH, dissolved organic carbon (DOC), CFC, 14C, hydrographic, and other chemical measurements.

The R/V Ronald H. Brown A16S_2005 data set is available free of charge as a numeric data package (NDP) from the Carbon Dioxide Information Analysis Center (CDIAC). The NDP consists of the oceanographic data files and this printed documentation, which describes the procedures and methods used to obtain the data.

Keywords: carbon dioxide, total CO2, total alkalinity, pH, fugacity of CO2, carbon cycle, coulometry, potentiometry, hydrographic measurements, CLIVAR, Atlantic Ocean

1. Background Information

The cruise of research vessel (R/V) Ronald H. Brown along section A16S from Punta Arenas, Chile, to Fortaleza, Brazil, was the second in the National Oceanic and Atmospheric Administration (NOAA) -led contributions to a series of repeat hydrography cruises to measure decadal changes in circulation, heat and fresh water budgets, and carbon inventory in the ocean. The cruises repeat a subset of the World Ocean Circulation Experiment Hydrographic Program (WHP) and Joint Global Ocean Flux Study (JGOFS) lines occupied in each major ocean basin in the 1990s.

The WOCE/WHP program is driven by the need to monitor the increases in carbon dioxide (CO2) in the ocean and provide the necessary data to support continuing model development that will lead to improved forecasting skills for oceans and global climate. During the 1990s, the WOCE/JGOFS survey provided a full-depth baseline data set against which to measure future changes. By integrating the scientific needs of programs requiring measurement of the full water column, major synergies and cost savings are achieved. These measurements are of importance both for major research programs, such as Climate Variability (CLIVAR) and the U.S. Global Climate Research Project (GCRP) Ocean Carbon and Climate Change Program (OCCC), and for operational activities such as the Global Ocean Observation System (GOOS) and the Global Climate Observing System (GCOS). As outlined in the program documentation, one component of a global observing system for the physical climate/CO2 system should include periodic observations of hydrographic variables, CO2 system parameters, and other tracers. The large-scale observation component of the OCCC program has a need for systematic observations of the invasion of anthropogenic carbon in the ocean that is superimposed on a variable natural background. The five topical areas addressed by the CO2/CLIVAR repeat hydrography program are:

  1. carbon system studies;
  2. heat and freshwater storage and flux studies;
  3. deep and shallow water mass and ventilation studies;
  4. calibration of autonomous sensors; and
  5. data for model calibration.

A hydrographic survey consisting of a meridional lowered acoustic Doppler current profiler (LADCP)/conductivity, temperature, and depth (CTD)/rosette section in the western South Atlantic was carried out in January - February 2005 (see also Wanninkhof and Doney, 2005 for additional cruise details). The R/V Ronald H. Brown departed Punta Arenas, Chile on 11 January 2005. Starting from 60°S, a total of 121 LADCP/CTD/Rosette stations were occupied (see Fig. 1.1) and 12 Argo floats and 11 drifters were deployed from 17 January - 21 February. Up to thirty-six water samples and continuous LADCP, CTD, and bio-optical data were collected on each cast to within 20 m of the bottom. Station spacing was nominally 0.5° of latitude except near the continental slope of South Georgia Island where a shorter station spacing was used to resolve boundary currents. Fewer than 36 discrete water samples were also collected on the shallow, continental slope stations as well as for some of the early stations due to heavy weather conditions. Salinity, dissolved oxygen, and nutrient samples were analyzed from every bottle sampled on the rosette. Other parameters from the bottles were sampled at a lower density. The cruise ended in Fortaleza, Brazil, on 24 February 2005.

This data report focuses on the measurements of total inorganic carbon or dissolved inorganic carbon (TCO2 or DIC), fugacity of CO2 (fCO2), total alkalinity (TALK), pH, dissolved organic carbon (DOC), chlorofluorocarbons (CFC-11 and CFC-12), 14C, nitrate (NO3), nitrite (NO2), phosphate (PO4), silicate (SiO4), and dissolved oxygen (O2).

The methodology, instrumentation, and standardization of these parameters improved significantly during the WOCE/JGOFS era. Notable developments include release of manuals detailing the analytical methods and operating protocols (DOE 1994). Certified reference materials (CRMs) are now available for TCO2 and TALK, which are run interspersed with samples to determine calibration offsets. For this cruise, the TALK and TCO2 values were adjusted to account for the small difference between the CRMs run at sea and the certified value determined at Scripps Institution of Oceanography (SIO). The TCO2 coulometers were calibrated daily by injecting aliquots of pure CO2 (99.995%). The stability of each coulometer cell solution was confirmed with the analyses of several CRMs each day.

image

Fig. 1. Cruise track for the Atlantic Ocean A16S_2005 cruise in January - February 2005.

Instrumentation has improved in the last decade. Alkalinity measurements can be done with better precision through automation and close checks of the response of electrodes. Burettes are independently calibrated, and the preparation of titrant (hydrochloric acid) has undergone improved quality control and standardization (Millero et al. 1998). Measurement of pH is now done with extreme precision with spectrophotometric methods (Byrne and Breland 1989). The TCO2 measurements are done by coulometry, a precise integrative method. During the A16S_2005 cruise two single-operator multiparameter metabolic analyzers (SOMMAs) (Johnson et al. 1999) were utilized for analyses, which facilitated a sample throughput of up to 80 samples per day. The fCO2 measurements were done with an equilibration system described in Wanninkhof and Thoning (1993). Oxygen measurements were performed by Winkler titrations (Carpenter 1965) with photometric endpoint detection (Friederich, Sherman, and Codispoti 1984).

The data underwent careful quality assurance and quality control (QA/QC) both during and after the cruise. The precision of the measurements was determined from duplicate sampling and comparison of data from deep water, where little variability is expected. Outliers in the data were flagged based on several methods utilizing prior knowledge of the trends and known relationships between parameters. Depth profiles for each parameter were scrutinized for outliers. When deviations were observed, other parameters were assessed to determine whether they showed deviations as well. Inorganic carbon system parameters were linked through physical and chemical properties. In particular, if two of the four carbon system parameters are measured, the other two can be calculated, provided that the silicate concentration, the phosphate concentration, the temperature, and the salinity of the sample are known. These so-called internal consistency calculations were used to assess the difference between calculated and measured values. When the difference between the measured TALK and the TALK calculated from TCO2 and pH or fCO2 exceeded 10 µmol/kg, the three parameters were scrutinized and compared with other parameters to assess whether the datum should be labeled as questionable. Other techniques, described in detail below, include regional multi-linear regressions (MLR) between the inorganic carbon parameters and physical and chemical parameters known to correlate with them [for instance, TCO2 = f(T, S, AOU, Si, PO4)]. Again, the differences between measured and calculated parameters were inspected. Finally, the parameters were plotted against latitude for narrow depth intervals. Since changes along depth horizons are usually gradual, anomalies can be easily spotted and flagged.

This report describes procedures and methods for hydrographic measurement and the analytical procedures, calculations, and assessment of precision for nutrient, oxygen, TCO2, TALK, fCO2, pH, CFCs, 14C, and DOC measurements. A description of the QA/QC methods based on internal consistency of these parameters and the MLR technique is also provided.

2. Description of the Expedition

2.1 R/V Ronald H. Brown: Technical Details and History

The NOAA ship R/V Ronald H. Brown, a state-of-the-art oceanographic and atmospheric research platform, is the largest vessel in the NOAA fleet. With its highly advanced instruments and sensors, R/V Ronald H. Brown travels worldwide supporting scientific studies to increase our understanding of the worlds oceans and climate. Commissioned on July 19, 1997, in its home port of Charleston, South Carolina, Ronald H. Brown has sailed in the Pacific, Atlantic, and Indian Oceans. The ship was named in honor of former Secretary of Commerce Ronald H. Brown, who was killed in a plane crash on April 3, 1996, while on a trade mission to Bosnia. R/V Ronald H. Brown is operated by NOAA Marine and Aviation Operations and carries a complement of 6 NOAA Corps officers, 20 crew members, and a maximum of 33 scientists. Table 2.1. provides a detailed description of the ship.

Table 2.1. Specifications of R/V Ronald H. Brown
Designer VT Halter Marine, Inc.
Builder VT Halter Marine, Inc., Moss Point, Mississippi
Launched May 30, 1996
Delivered April 18, 1997
Commissioned July, 19, 1997
Hull number R104
Call letters WTEC
Home port Charleston, South Carolina
Length 83.5 m (274 ft)
Breadth (molded) 16.0 m (52.5 ft)
Draft, maximum 5.2 m (17.0 ft)
Depth to main deck 8 m (26.5 ft)
Displacement 3,250 tons
Maximum speed 15 kn (9 kn on one generator)
Cruise speed 12 kn
Range 11,300 nm at 12-kn speed plus 30 days on station
Total crew and scientists 58
Maximum cruise duration 60 days (45 day practical limit)
Science quarters Main lab: 1,730 ft2
Electronics/computer lab: 720 ft2
Wet lab: 230 ft2
Hydro lab: 700 ft2
Biochemical lab: 720 ft2

2.2 R/V Ronald H. Brown A16S_2005 Cruise Information
Ship name Ronald H. Brown
EXPOCODE 33RO200501
CLIVAR section A16S_2005
Ports of call Punta Arenas, Chile; Fortaleza, Brazil
Dates January 11 - February 24, 2005
Funding support NOAA, NSF
Chief scientists Dr. Rik Wanninkhof, NOAA/AOML
Dr. Scot Doney, WHOI

2.3 Parameters Measured, Institution, and Responsible Investigators
Parameter Institution Responsible Investigator
CTD PMEL/AOML G. Johnson/M. Baringer
ADCP/LADCP UH/LDEO E. Firing/A. Thurnherr
Salinity PMEL G. Johnson
Nutrients UW/AOML C. Mordy, J.-Z. Zhang
Dissolved Oxygen RSMAS-UM C. Langdon
CFCs PMEL, UW M. Warner, J. Bullister
Tritium, Helium LDEO P. Schlosser
TCO2 AOML, PMEL R. Wanninkhof, R. Feely
Discrete pCO2 AOML R. Wanninkhof
TALK, pH RSMAS/UM F. Millero
DOC RSMAS/UM D. Hansell
CDOM UCSB N. Nelson/C. Carlson
Underway pCO2 AOML R. Wanninkhof
CO2 System Develop. USF R. Byrne
14C, 13C WHOI A. McNichol

2.4 Participating Institutions
PMEL Pacific Marine Environmental Laboratory
AOML Atlantic Oceanographic and Meteorological Laboratory
LDEO Lamont-Doherty Earth Observatory
RSMAS/UM Rosenstiel School of Marine and Atmospheric Science, University of Miami
UCSB University of California at Santa Barbara
UH University of Hawaii
USF University of South Florida
WHOI Woods Hole Oceanographic Institution
UW University of Washington

3. Description of Variables and Methods

3.1 Hydrographic Measurements

Samples for CFCs, helium isotopes (3He), oxygen (O2), hydrochlorofluorocarbon (HCFCs), partial pressure of CO2 (pCO2), TCO2, hydrogen ion activities (pH), TALK, radiocarbon (Δ14C), tritium, DOC, chromophoric dissolved organic matter (CDOM), particulate inorganic/organic carbon (PIC/POC), salinity, and nutrients were drawn in this sequence from a CTD sampling package containing thirty-six 12-L Bullister bottles. A detailed descriptions of methods for the CTD data, LADCP data and biooptical data are given in Wanninkhof and Doney (2005). Oxygen, nutrient, and salinity samples were taken from all bottles. Oxygen draw temperature readings were commenced after station 25. For the other parameters, not all stations or all bottles were sampled. The stations at full degrees of latitude (odd numbered stations) were generally completely sampled for CFCs, TCO2, pH, and TALK, with partial sampling for DOC and CDOM. The half-degree stations were partially sampled for HCFC, PIC/POC, CFCs, TCO2, pH, and TALK. Discrete pCO2 profiles were obtained at every two degrees. 3He, Δ14C, and tritium were sampled at different intervals (primarily at full-latitude stations).

A total of 4174 bottle salinity samples were taken during the cruise, of which 127 were flagged as questionable and 3 as bad. Two samples were lost during analysis. Samples were drawn from the 12-L Bullister bottles into 250-mL Kimax borosilicate bottles. The bottles were rinsed at least three times before being filled to approximately 220 mL. A plastic insert and Nalgene cap were used to seal the sample in the bottle. At the conclusion of sampling, the time was noted and samples were placed into the salinometer lab so they could equilibrate to room temperature. Samples were analyzed after a period of at least 10 hours and typically not more than 24 hours from the time of sampling. Samples were run on a Guildline 8400B Laboratory Salinometer, serial number 60843. The salinometer had been last calibrated at Guildline in January of 2004. IAPSO Standard Seawater was used to standardize the instrument. The software used (ASALW) was developed at Scripps Institution of Oceanography. As per the instructions provided in the software, the cell was rinsed at least two times with sample at a relatively fast flow rate; the flow was adjusted to a slower rate for the final fill, and a reading was taken. The cell was drained and slowly filled for a second reading. If the two readings agreed within 0.00005, the values were accepted; otherwise, an additional reading was required. PSS-78 salinity (UNESCO, 1981) was calculated. Corrections were applied to the data for differences between beginning and ending standards.

Samples for dissolved oxygen analyses were drawn from 12-L Bullister bottles into calibrated 140-mL iodine titration flasks using Tygon tubing with silicone adapters that fit over the petcock to avoid contamination of DOC samples. Bottles were rinsed twice and filled from the bottom, then overflowed three volumes (care was taken to avoid entraining any bubbles). One-mL of MnCl2 and one-mL of NaOH/NaI were added, then the flask was stoppered and shaken. Deionized water was added to the neck of each flask to create a water seal. The flasks were stored in the lab in plastic totes at room temperature for 12 hours before analysis. A total of 4659 samples were analyzed during the cruise, of which 37 samples were flagged as questionable and 4 as bad. Three samples were not reported. The whole-bottle titration technique of Carpenter (1965) was performed with modifications by Culberson et al. (1991), but with a more dilute solution of thiosulfate (10 g/L). Dissolved oxygen analyses were performed with a Monterey Bay Aquarium Research Institute (MBARI)-designed automated oxygen titrator using photometric end-point detection based on the absorption of 365-nm wavelength ultra-violet light. The titration of the samples and the data logging were controlled by a 386 PC running the oxygen program written by Gernot Friedrich (Friederich et al. 1984). Thiosulfate was dispensed by a Dosimat 665 fitted with a 5.0-mL burette. The autotitrator and Dosimat performed well.

Dissolved nutrient (phosphate, silicic acid, nitrate, and nitrite) samples were drawn in 40-mL HDPE Boston Round sample bottles that had been stored in 10% HCl and rinsed four to five times with sample before filling. A replicate was always drawn from the deep bottle for analysis on the subsequent station. All samples were brought to room temperature prior to analysis. A separate analytical run was conducted at each station (except for the shallowest stations). An analytical run consisted of blanks and working standards, old working standards, deep water from the previous station, samples analyzed from deep to surface, replicate analysis of the four deep samples and any problem samples, and finally the working standards and blanks. The blanks were deionized water, and the standards were a zero standard in Low Nutrient Seawater (LNSW) and a high standard. Linearity of the autoanalyzer was checked every ten days, and corrections for non-linearity were applied during final data reduction.

All nutrients were measured using automated continuous flow analysis with a segmented flow and colorimetric detection. The four-channel autoanalyzer was customized using components from various systems. The major components were an Alpkem 301 sampler, two 24-channel Ismatek peristaltic pumps, four Thermo-Separation monochrometers, and custom software for digitally logging and processing the chromatographs. Glass coils and tubing from the Technicon Autoanalyzer II were used for analysis of phosphate, and micro-coils from Alpkem were used for the other three analyses. The detailed methods are described by Gordon et al. (1993). Because pump tubing destined for the cruise was lost in transit, some of the pump tube sizes suggested in the manual had to be modified. Pump tubes were changed four times during the expedition.

Silicic acid was analyzed using a modification of Armstrong et al. (1967). An acidic solution of ammonium molybdate was added to a seawater sample to produce silicomolybic acid. Oxalic acid was added to inhibit a secondary reaction with phosphate. Finally, the reduction with ascorbic acid formed the blue compound silicomolybdous acid. The color formation was detected using a 6-mm flowcell at 660 nm. The use of oxalic acid and ascorbic acid (instead of tartaric acid and stannous chloride as suggested by Gordon et al.) was to reduce the toxicity of our waste stream.

Nitrate and nitrite analyses were also modified from Armstrong et al. (1967). Nitrate was reduced to nitrite in a cadmium column, then formed into a red azo dye by complexing nitrite with sulfanilamide and N-1-naphthylethylenediamine. The color formation was detected using a 6-mm flow cell at 540 nm. The same technique was used to measure nitrite (excluding the reduction step), but the color formation was detected using a 10-mm flow cell at 540 nm.

Phosphate analysis was based on the technique of Bernhardt and Wilhelms (1967). An acidic solution of ammonium molybdate was added to the sample to produce phosphomolybdic acid, and this was reduced to the blue compound phosphomolybdous acid following the addition of hydrazine sulfate. The reaction was heated to 55°C to bring the reaction to completion, and color formation was detected using a 10-mm flow cell at 815 nm.

3.2 Total CO2 Measurements

Samples for TCO2 measurements were drawn according to procedures outlined in the Handbook of Methods for CO2 Analysis (DOE 1994) from 12-L Bullister bottles into 540-mL Pyrex bottles using Tygon tubing with a silicone adapter on the petcock to avoid contamination of DOC samples. Bottles were rinsed and filled from the bottom, leaving 5 mL of headspace; care was taken not to entrain any bubbles. After 0.2 mL of saturated HgCl2 solution was added as a preservative, the sample bottles were sealed with glass stoppers lightly covered with Apiezon-L grease and were stored at room temperature for a maximum of 12 hours prior to analysis.

TCO2 samples were collected at every degree from all depths (typically 36 for standard deep water stations) with three replicate samples. Some samples were also collected at every half-degree. The replicate seawater samples were taken from the surface, 1000 m, and bottom Bullister bottles and run at different times during the cell. The first replicate of the bottom water was used at the start of the cell with fresh coulometer solution, and the first of the 1000-m replicates was run in the middle of the cell after about 12 mg of C were titrated. The second one of the bottom replicates was run at the end of the cell after about 25 mg of C were titrated. A new coulometer cell was started with the second of the 1000-m replicates and the first of the surface replicates. In the middle of this cell, the second of the surface replicates was run and the first one of the surface duplicates of a partial station. The second of the partial station duplicates was run at the end of this cell. No systematic difference between the replicates was observed. There was no systematic dependency of results with an amount of carbon titrated for a particular cell. A total of 2482 samples for TCO2 were collected and analyzed during the cruise.

The TCO2 analytical equipment was set up in a seagoing laboratory van. The analysis was done by coulometry with two analytical systems (called AOML-1 and AOML-2) used simultaneously on the cruise. Each system consisted of a coulometer (UIC, Inc.) coupled with a single operator multi-parameter metabolic analyzer (SOMMA) inlet system developed by Kenneth Johnson (Johnson et al. 1985, 1987, 1993; Johnson 1992) now retired from Brookhaven National Laboratory (BNL). In the coulometric analysis of TCO2, all carbonate species are converted to CO2 (gas) by addition of excess hydrogen ion (acid) to the seawater sample, and the evolved CO2 gas is swept into the titration cell of the coulometer with pure air or compressed nitrogen, where it reacts quantitatively with a proprietary reagent based on ethanolamine to generate hydrogen ions. In this process, the solution changes from blue to colorless, which triggers a current through the cell and causes coulometrical generation of OH ions at the anode. The OH ions react with the H+, and the solution turns blue again. A beam of light is shone through the solution, and a photometric detector at the opposite side of the cell senses the change in transmission. Once the percent transmission reaches its original value, the coulometric titration is stopped, and the amount of CO2 that enters the cell is determined by integrating the total charge during the titration.

The coulometers were calibrated by injecting aliquots of pure CO2 (99.995%) by means of an 8-port valve outfitted with two sample loops with known gas volumes (AOML-1: 1.9951 mL at 25.05°C and 0.9807 mL at 25.10°C; AOML-2: 2.0018 mL at 25.09°C and 0.9949 mL at 25.06°C) bracketing the amount of CO2 extracted from the water samples for the two AOML systems.

The stability of each coulometer cell solution was confirmed three different ways: the Certified Reference Material (CRM), Batch 66, supplied by Dr. A. Dickson of SIO, was measured at the beginning and the middle, gas loops in the beginning and at the end, and the duplicate samples at the beginning, middle, and end of each cell solution (Fig 3.1., Table3.1.). The coulometer cell solution was replaced after 25 mg of carbon was titrated, typically after 912 hours of continuous use.

The pipette volume was determined prior to the cruise by taking aliquots at known temperature of distilled water from the volumes. The weights with the appropriate densities were used to determine the volume of the pipettes (AOML1: 28.716 mL at 20.00°C, AOML2: 22.547 mL at 20.00°C).

Calculation of the amount of CO2 injected was according to the CO2 handbook (DOE 1994). The concentration of CO2 ([CO2]) in the samples was determined according to:

image

where Cal. Factor is the calibration factor, Counts is the instrument reading at the end of the analysis, Blank is the counts/minute determined from blank runs performed at least once for each cell solution, Run Time is the length of coulometric titration (in minutes), and K is the conversion factor from counts to mol.

The instrument has a salinity sensor, but all TCO2 values were recalculated to a molar weight (µmol/kg) using density obtained from the CTDs salinity sensor. The TCO2 values were corrected for dilution by 0.2 mL of HgCl2 used for sample preservation. The total water volume of the sample bottles was 540 mL. The correction factor used for dilution was 1.00037. A correction was also applied for the offset from the CRM. This correction was applied for each cell using the CRM value obtained in the beginning of the cell. The results underwent initial quality control on the ship using property plots: TCO2 vs.Depth, TCO2 vs. Potential Temperature, TCO2 vs. AOU, TCO2 vs. NO3; TCO2 vs. SiO3, TCO2 vs. PO4, TCO2 vs. TALK, and TCO2 vs. pH. Also contour plots of TCO2 vs. LAT and Depthwere used to analyze the quality of the data.

image

Fig. 3.1. Results of the duplicate TCO2 samples collected during the R/V Ronald H. Brown cruise along the Atlantic Ocean section A16S_2005.

Table 3.1. TCO2 duplicate measurement statistics
Measurement method Avg. Std. dev. No.
Duplicate samples measured back-to-back 0.9 0.68 64
One duplicate measured at the beginning; one at the middle of the cell 1.0 0.82 62
One duplicate measured at the middle; one at the end of the cell 1.1 0.91 34
One duplicate measured at the beginning; one at the end of the cell 1.3 1.02 72
Duplicates run on the same instrument but on different cells 1.2 0.83 102
Duplicates run on different instruments     0
Duplicates measured at the beginning of the cell, but not back-to-back 1.3 0.89 6
Duplicates measured at the middle of the cell, but not back-to-back 0.6 0.63 12
Duplicates measured at the end of the cell, but not back-to-back 0.7 0.98 4
Total no. of measurements 356

The overall performance of the instruments was good during the cruise. The air purifier supplying carrier and pneumatic gas malfunctioned at station 24. Compressed tanks of ultra-high purity nitrogen gas were used thereon. At the same time, soda lime traps used to scrub any CO2 from the carrier gas were removed from the air/N2 line, since they developed cracks over time and also appeared to release CO2 in pulses into the carrier. A coulometer was replaced during the test cast runs. It did not find an endpoint and did not stop counting. A number of pinch valves failed and were replaced; some cell caps began to leak, and some electrode leads broke. Finally, the Orbo tubes (filled with silica gel to absorb possible acid vapors) tended to break and leak and were not used after station 109 on either system.

Due to concerns about the large amount of water used for a TCO2 sample and use of grease on the stoppers that could contaminate samples for dissolved organic matter (DOM), comparison tests were performed with samples drawn in 250-mL borosilicate bottles with ground glass stoppers stored under cold water with the regular sampling procedures outlined above. The results are shown in Table 3.2.

Table 3.2. Test results of different sample bottle sizes for TCO2 measurements
Type Bottle2 RT3 TCO2 µmol/kg Bullister Bottle No Volume (mL) Average µmol/kg
CRM1 213 14 1971.11      
bottle A9 13 2122.29 9 500  
bottle C9 13 2121.10 9 500 2121.69
bottle S76 10 2119.95 9 250  
bottle S77 12 2120.01 9 250 2119.98
bottle S78 12 2120.28 10 250  
bottle S79 11 2120.89 10 250 2120.58
bottle A10 13 2121.04 10 500  
bottle C10 12 2122.5 10 500 2121.77
bottle A11 12 2121.37 11 500  
bottle C11 14 2123.05 11 500 2122.21
CRM 213 15 1972.10      
bottle S80 13 2120.74 11 250  
bottle S81 19 2121.09 11 250 2120.915
bottle S82 11 2120.51 12 250  
bottle S83 17 2121.03 12 250 2120.77
bottle A12 20 2126.01 12 500  
bottle C12 20 2125.5 12 500  
bottle A13 20 2128.34 13 500  
bottle C13 17 2122.12 13 500 2122.12
bottle S84 11 2120.61 13 250  
bottle S85 11 2119.34 13 250 2119.97
bottle S86 9 2119.26 14 250  
bottle S87 9 2119.41 14 250 2119.33
bottle A14 9 2120.74 14 500  
bottle C14 9 2121.61 14 500 2121.17

1CRM = 1969.57 µmol/kg;
2Bottles with prefix S are the 250-mL bottles; the others are the 540-mL bottles used during the cruise;
3RT: run time of the sample on the coulometer in minutes.
4Average: Average of duplicate large or duplicate small bottles taken from a particular Bullister bottle.

These tests and more ad-hoc tests at the beginning of the cruise showed a small but systematic difference between the small and large sample bottles: the average for large bottles was 2121.79 µmol/kg; for small bottles, 2120.26 µmol/kg; standard deviations were 0.37 and 0.55, respectively. The cause of the artifact is not clear, but the tests led to the decision to use the 500-mL greased stopper bottles for the entire cruise.

3.3 Discrete Fugacity of CO2 Measurements

Samples for fCO2 were drawn from 12-L Bullister bottles into 500-mL Pyrex volumetric flasks using Tygon tubing with a Silicone adapter over the petcock to avoid contamination of DOC samples. Bottles were rinsed twice, filled from the bottom, and overflowed half a volume while care was taken that no bubbles were entrained. About 5 mL of water was withdrawn by removing the pinched-off sampling tube from the neck of the flask to create a small expansion volume, then 0.2 mL of saturated mercuric chloride (HgCl2) solution was added as a preservative. The sample bottles were sealed with a polyethylene-lined screw cap. The samples were stored upside down in coolers at room temperature for a maximum of 10 hours.

Thirty samples were drawn at every fourth station (at 2 intervals) for a total of 847 samples from 29 stations. In addition, samples were drawn in duplicate from the underway seawater line at 6-hour intervals between Punta Arenas and the start of the line at 60°S, 31°W.

The fCO2 was measured on the A16S_2005 cruise at a constant temperature of 20°C by equilibrating a 500-mL water aliquot in a volumetric flask with a closed headspace. The headspace was circulated through a nondispersive infrared detector that measures both CO2 and H2O levels. The analytical instrumentation is detailed in Wanninkhof and Thoning (1993) and is similar to the setup used in the S.Atl-91 cruise that occupied the same cruise line from 43°S northward in 1991 (Forde et al. 1994). Discrete fCO2 measurements were also made on the SAVE-5/HYDROS-4 cruises using a system developed by Chipman et al. (1993) that covered a similar track from 54°S to 2°N in 1989-1990 (SAVE 1992).

In short, in the system used on A16S_2005, a 500-mL water sample was equilibrated at ambient pressure with an 80-mL headspace in a volumetric flask held at a temperature of 20 °C in a thermostatted bath. The headspace was circulated through a nondispersive infrared analyzer (NDIR), LICOR, model 6262. Upon equilibration, the circulation flow was stopped and 20 consecutive 1-second readings of H2O content and CO2 content in the cell were taken and averaged. The system was a dual-channel system where one equilibration occurs while circulating through the NDIR and a second flask was equilibrated offline. Once the first sample was analyzed, the second flask was switched in line with the NDIR and the residual air in the NDIR was equilibrated with the second flask content. The second equilibration phase through the NDIR took less time as a large part of the headspace was already equilibrated offline. The two-channel configuration decreases the total analysis time to about 20 minutes for two samples. An expandable volume in the circulation loop near the flasks consisting of a small deflated balloon kept the content of flasks at room pressure.

To account for instrument drift and to maintain measurement precision, a set of six gas standards was run through the system before and after every eight seawater samples. The standards of CO2 in natural air were obtained from Scott-Marin and referenced against primary standards purchased from C.D. Keeling in 1991. The primary standards were on the WMO-78 scale. The mole fractions of the standards, which were run in the following sequence, were:

Tank number Mole fraction
CA05989 378.71 ppm
CA05980 792.51 ppm
CA05984 1036.92 ppm
CA05940 1533.70 ppm
CA05988 593.64 ppm
CA05998 205.07 ppm

The standards were also used as the headspace gas for the equilibration. Since the mole fractions of the gases in the headspace prior to equilibration were known, the small perturbation of the fCO2 in the water during the equilibration process could be quantitatively accounted for. The headspace gas was selected so that it is close to the anticipated water value, thereby minimizing the correction.

The determination of fCO2 @20°C [fCO2(20)] in water from the headspace measurement involved several steps. The infra-red (IR) detector response for the standards was normalized for temperature. The IR analyzer output of dry mole fraction of CO2 (xCO2) for samples was normalized to 1 atm pressure. The sample values were converted to the true mixing ratio based on a second-order polynomial fit between the instrument xCO2 readout and the values of the three nearest concentrations of the compressed gas standards. The mixing ratio in the headspace was converted to a fugacity assuming 100% humidity and corrected to fugacity of CO2 in the water sample prior to equilibration by accounting for change in total CO2 in water during the equilibration process (for details see Wanninkhof and Thoning 1993). The change in fCO2(20) caused by the change in TCO2 was calculated using the constraint that TALK remains constant during exchange of CO2 gas between the headspace and the water. The calculation is outlined Peng et al. (1987).

Uncertainty based on duplicate sampling of the same Bullister bottle for fCO2 analysis was determined on select stations of the cruise. The comparisons are presented In Table 3.3

Table 3.3. Comparison of duplicate discrete fCO2 samples.
Station Sample No. fCO2av fCO2 % difference
5 203 1093.8 3.0 0.3
5 209 1089.0 2.3 0.2
9 103 1087.3 1.5 0.1
9 105 1088.8 2.4 0.2
9 109 1081.6 2.7 0.2
21 135 572.7 0.4 0.07
49 121 838.0 0.4 0.05
65 121 756.1 5.6 0.7
93 124 1040.3 2.3 0.2

fCO2av = average of the duplicate samples.
fCO2 = absolute difference between the duplicates
% difference = ∆fCO2/fCO2av X 100

Measuring TCO2, TALK, and fCO2(20) on the cruise, offers an opportunity to assess the internal consistency of these inorganic carbon parameters. For the analyses, the apparent carbonate dissociation constants determined by Mehrbach (1973) as refitted by Dickson and Millero (1987) were used. An Excel program developed by D. Pierrot of RSMAS/University of Miami using the code developed by Lewis and Wallace (1998) facilitated the calculations. The average difference between measured alkalinity and that calculated using TCO2 and fCO2(20) was -2.03 ±4.91 µmol/kg (n=768). The agreement was the best obtained to date and suggested an excellent data quality of these parameters. A compilation of the difference plotted versus depth is shown in Fig. 3.2.

image

Fig. 3.2. Difference between measured TALK and that calculated from TCO2 and fCO2(20) plotted versus depth.

The distribution of differences is presented in Fig. 3.3. The figure shows a near-normal distribution with a slight negative bias suggesting predominantly random errors.

image

Fig. 3.3. Distribution of measured TALK – calculated TALK from TCO2 and fCO2(20).

The A16S_2005 cruise overlapped with the A16N_2003a cruise from 6˚S to 2.5˚S and measurements of fCO2(20) were performed at 4˚S and 6˚S on both cruises. The depth profiles shown in Figs. 3.4 and 3.5 show excellent agreement between the observations at the two locations. For example, for the depth interval between 1800 and 3500 dB in the core of the North Atlantic deep water where fCO2(20) is quite constant, the average of the values of the two overlap stations are well within one standard deviation (9 atm) of each other with no significant differences between the 2003 and 2005 data.

image

Fig. 3.4. Comparison of fCO2(20) data at 6°S for 2003 and 2005.

image

Fig. 3.5. Comparison of fCO2(20) data at 4°S for 2003 and 2005.

3.4 Total Alkalinity Measurements

Samples for total alkalinity measurements were drawn from 12-L Bullister bottles into 500-mL borosilicate flasks using silicone tubing fit over the petcock to avoid contamination of DOC samples. Bottles were rinsed a minimum of two times and filled from the bottom and a quarter of a volume was overflowed while taking care to avoid the entrainment of any bubbles. Approximately 15 mL of water was withdrawn from the flask by arresting the sample flow and removing the sampling tube, thus creating a small expansion volume and a reproducible headspace. The sample bottles were sealed at a ground glass joint with a glass stopper. The samples were stored at room temperature for a maximum of 7 hours. Thirty-six samples and three duplicates were drawn on odd number stations (at 1 intervals) for a total of 61 full stations. Typically, 9 depths were sampled with a duplicate at the surface at the 60 half stations. Periodically, multiple duplicate samples were drawn with a specific focus on photic zone and region of high DOC. The purpose was to determine the difference in TALK after filtration with a 0.45-m nylon membrane filter. Additional underway samples were drawn in duplicate from the underway seawater line at 6-hour intervals between Punta Arenas, Chile, and the start of the line at 60°S, 31°W

The TALK was evaluated from the proton balance at the alkalinity equivalence point, pHequiv = 4.5 at 25°C and zero ionic strength in 1 kg of sample. The method utilizes a multi-point hydrochloric acid titration of seawater according to the definition of total alkalinity (Dickson 1981).

The titration system consisted of a Metrohm 665 Dosimat titrator, an Orion 720A pH meter, and a custom-designed, plastic water-jacketed titration cell (Millero et al. 1993a). Both the seawater sample and acid titrant are temperature-equilibrated to a constant temperature of 25 ± 0.1°C with a water bath (Neslab, model RTE-17). The plastic water-jacketed cell is similar to the cells used by Bradshaw and Brewer (1988) except a larger volume (~200 mL) is employed to increase the precision. Each cell has a fill-and-drain valve, which increases the reproducibility of the volume of sample contained in the cell. The titration acidified seawater passed the carbonic acid endpoint by adding HCl stepwise through an injection tip into the cell. A typical titration recorded the (electro-magnetic field) EMF after the readings became stable (deviation less than 0.09 mV), and then enough acid was added to change the voltage a preassigned increment (13 mV). A full titration (~25 points) takes about 20 min. The electrodes used to measure the EMF of the sample during a titration consisted of a ROSS glass pH electrode (Orion, model 810100) and a double-junction Ag, AgCl reference electrode (Orion, model 900200).

A single 50-L batch of ~0.25-m HCl acid was prepared in 0.45-m NaCl by diluting concentrated HCl, AR Select Mallinckrodt, to yield a total ionic strength similar to seawater of salinity 35.0 (I ≈ 0.7 M). The acid was standardized by a coulometric technique (Marinenko and Taylor 1968, Taylor and Smith 1959) and verified with alkalinity titrations on seawater of known alkalinity. Furthermore, Andrew Dicksons laboratory performed an independent determination of the acid molality on sub-samples. The calibrated molality of the acid used was 0.2434 ± 0.0001 m HCl. The acid was stored in 500-mL glass bottles sealed with Apiezon L grease for use in the field.

The volumes of the cells used were determined to ± 0.03 mL in the laboratory by multiple titrations using seawater of known total alkalinity and CRM. Calibrations of the burette of the Dosimat with water at 25°C indicate that the systems deliver 3.000 mL (the approximate value for a titration of seawater) to a precision of ± 0.0004 mL, resulting in an error of ± 0.3 µmol/kg in TALK and TCO2. The reproducibility and precision of measurements are checked using low nutrient surface seawater and CRM, Batch Nos. 59 and 66. CRMs were utilized to account for instrument drift and to maintain measurement precision. Duplicate analyses provide additional QA and were taken from same Bullister bottle.

The assigned values of the CRM provided by A. Dickson of SIO were:

  • Batch #59: Total alkalinity: 2220.98 ± 0.58 µmol/kg Salinity: 33.316
  • Batch #66: Total alkalinity: 2193.27 ± 0.60 µmol/kg Salinity: 32.962

An integrated program controlled the titration, data collection, and the calculation of the carbonate parameters (TALK, pH, and TCO2) (Millero et al. 1993b). The program is patterned after those developed by Dickson (1981), Johansson and Wedborg (1982), and DOE (1994). The program uses a Levenberg-Marquardt nonlinear least-squares algorithm to calculate the TALK and TCO2 from the potentiometric titration data.

CRM Instrument 1 Instrument 2
Total Number of Sets 56 52
Number of Sets Used 48 46
Standard Deviation ± 3.5 µmol/kg

Duplicates Between Systems Instrument 1 Instrument 2
Total Number of Sets 143 31 42
Number of Sets Used 130 25 35
Standatd Deviation ± 2.9 µmol/kg ± 1.6 µmol/kg ± 1.3 µmol/kg

Note: Duplicate samples with differences three times larger than standard deviation were omitted from the analyses. The number omitted is the difference between the total number of sets and the sets used.

At the beginning of the cruise, a titration cell was swapped out for a spare cell because of a combination of instability in the electrodes and an air bubble consistently being trapped. One valve and a proximity switch were replaced without discernible downtime. Sporadically, a solenoid valve at the bottom of the titration cell would fail to engage or disengage, resulting in the loss of the sample or a failed titration due to a poor rinse or an air bubble. The titration cell on system showed some drift in CRM values on February 4, 2005; the titration cells recalibration values were used to correct this. Communication problems between the software and the components of the TALK system were remedied with the replacement of cables and/or components. A Metrohm 665 Dosimat titrator and an Orion 720A pH meter were replaced. Computer instability resulted in the loss of one sample.

3.5 pH Measurements

Samples for pH measurements were drawn from 12-L Bullister bottles into 50-mL glass syringes using polycarbonate Luer-lock valves that fit in the petcock. Syringes were rinsed a minimum of three times and filled while care was taken to avoid the entrainment of any bubbles. A rubber band ensured positive pressure on the barrel of the syringe. The samples were stored at room temperature for a maximum of 7 hours. Thirty-six samples and 3 duplicates were drawn on odd- numbered stations (at 1 intervals) for a total of 61 full stations. At even-numbered stations, surface water and a duplicate were always taken; in addition, 5 to 20 other depths were also sampled, for a total of 60 half stations. Typically, nine depths were sampled with a duplicate at the surface for the half-stations. Underway samples were drawn in duplicate from the underway seawater line at 6-hour intervals between Punta Arenas, Chile, and the start of the line at 60°S, 31°W.

Measurements of the pH of seawater on the total hydrogen ion concentration pH scale (pHt) were performed using the multi-wavelength spectrophotometric techniques of Clayton and Byrne (1993). Determination of the absorbance at several wavelengths eliminates the need to know the concentrations of indicator in the sample. Sulphonephthalein indicators such as m-cresol purple (mCP), thymol blue, and cresol red are suitable for determining pH. The system is patterned after the standard operating procedure developed by DOE (1994) and utilizes mCP. This fully automated system performs discrete analysis of pH samples approximately every 12 minutes on a sample volume of 25 mL. A microprocessor-controlled syringe and sampling valve aspirates and injects the seawater sample into the 10-cm optical cell at a precisely controlled rate. The syringe rinses and primes the optical cell with 20 mL of sample and the software permits 5 minutes for temperature stabilization. A refrigerated circulating temperature bath (Neslab, model RTE-17) regulates the temperature of the sample at 25 ± 0.01°C. An Agilent 8453 UV/VIS spectrophotometer measures background absorbance of the sample. The automated syringe and sampling valves aspirates 4.90 mL seawater and 0.008 mL of indicator and injects the mixture into the cell. After the software permits 5 min for temperature stabilization, a Guildline 9540 digital platinum resistance thermometer measures the temperature and the spectrophotometer acquires the absorbance at 434, 578, and 730 nm.

A concentrated solution of 2.0 mM of mCP (C21H18O3S) dye of known pHt = 7.91 and R = 1.625 at 25°C were used.

A precision of better than 0.001 pH is possible with care, specifically with regard to temperature equilibration and sample handling. Measurements made on duplicate samples, TRIS buffers, and CRM, Batch No. 59, validate the precision and accuracy. Duplicate analyses provide additional QA and were taken from same Bullister bottles. The pHsws for the CRM was determined by spectrophotometric methods independently in the laboratory at RMSAS, University of Miami:

  • Batch #59: pHsws @ 25°C 7.9048 ± 0.0007 (n = 19)
  • Salinity 33.316

The pHt of the sample is perturbed by the addition of the indicator. The magnitude of this perturbation is a function of the difference between the seawater and indicator acidity. A correction factor applied for each batch of dye adjusts for this perturbation. For a 4.90-mL sample of seawater, 0.008 mL of mCP is added and the absorbance ratio is measured. From a second addition of mCP and a second absorbance ratio measurement, a change in the absorbance ratio per mL of added indicator is calculated. The value of the absorbance ratio (Rm) measured subsequent to the initial addition of the indicator was used to calculate R from:

R = Rm + (0.00095 – 0.00133 Rm) Vind (1)

where Vind is the volume of mCP used. Clayton and Byrne (1993) calibrated the mCP indicator using TRIS buffers (Ramette et al. 1977) and the equations of Dickson (1993). These equations are used to calculate pHt, the total scale in units of moles per kilogram of solution. The conversion of the pHt (µmol/kgH2O) to the seawater scale (µmol/kgsol) can be made using equations of Dickson and Millero (1987), Dickson and Riley (1979), and Dickson (1990).

CRM

Total number of sets: 136

Number of sets used: 124

CRM Batch #59: 7.9050 ± 0.0024 (pHsws at 25°C)

TRIS buffer

Total number of sets: 296

Number of sets used: 264

TRIS Buffer (0.04 m): 8.0935 ± 0.0019 (pHt @ 25°C)

Duplicates

Total number of sets: 291

Number of Sets used: 214

Standard deviation: ± 0.0019 (pHsws @ 25°C)

Note: The instrumental software automatically runs a duplicate analysis when the baseline absorbance at 730 nm is beyond a set threshold, thus a large number of duplicate results were omitted. Duplicate samples with differences more than three times larger than the standard deviation were omitted from the analyses. The number omitted is the difference between the total number of sets and the sets used.

At the start of the cruise, the outflow from the optical cell leaked into the thermostat-equipped water jacket; this was repaired by replacing the tubing. Sporadically, samples drawn from the syringe entrained an air bubble because the valve was improperly opened, the tubing was pinched, or the syringe plunger was dry and became stuck in the barrel. Some syringes suffered from fatigue at the metal Luer-lock and this resulted in the sample being lost or the analysis failing. Occasionally the software lost communication with the microprocessor-controlled syringe pumps and paused analysis; the problem was resolved by following the steps outlined in the software to reestablish communication.

3.6 Dissolved Organic Carbon Measurements

Water samples were collected from the rosette. Samples collected from the surface to 250 meters were filtered using precombusted (500°C) GF/F inline filters as they were being collected from the Niskin bottle. At depths > 250 meters, the samples were collected without filtration. After collection, samples were frozen upright in 60-mL acid-cleaned high-density polypropylene bottles and remained cold until analysis. Prior to analysis, samples were returned to room temperature then acidified to pH < 2 with concentrated hydrochloric acid. Analysis was performed onshore using a Shimadzu TOC-VCSH Total Organic Carbon Analyzer with the TNM-1 Total Nitrogen detector attached. Instrument conditions were as follows:

Combustion temperature - 680°C
Carrier gas - UHP Oxygen
Carrier flow rate - 150 mL/min
Ozone generation gas - Zero Air from Whatman TOC Gas Generator
Ozone flow rate - 500 mL/min
Sample sparge time - 2.0 minutes
Minimum number of injections - 3
Maximum number of injections - 5
Number of washes - 2
CV maximum (within sample) - 2.00%
Injection volume - 100 µL

The TOC system was calibrated using potassium hydrogen phthalate in Milli-Q water; the TN system was calibrated using potassium nitrate in Milli-Q water. System performance was verified daily using Consensus Reference Water distributed by the Hansell Organic Biogeochemistry Laboratory. This reference water is from deep in the Sargasso Sea that has been acidified and sealed in 10-mL ampoules, the concentration of which (44 µmol/kg C) has been determined by the consensus of up to six expert and independent laboratories. After verifying proper operation of the TOC/TN instrument, samples were set up on an auto sampler for analysis. The run started with a QW (Q Water) blank and a reference seawater analysis. Then six samples would be analyzed followed by another pair of QW blank and reference seawater analyses. This sequence would be repeated until all samples for that run were analyzed. The run ended with a QW blank, reference water, and a non-acidified QW blank. This was done to verify that the hydrochloric acid used to acidify the samples was uncontaminated. QW blanks and reference water samples were used to evaluate system performance during the analytical run. If a problem was detected with the blanks or reference waters, the samples were reanalyzed. The repeated use of deep sea consensus reference water in the analysis of DOC samples from this cruise indicates a between sample precision of approximately 0.6 µmol/kg, or a coefficient of variation of about 1.5%.

Chlorofluorocarbon Measurements

Samples for the analysis of dissolved CFC-11, -12, -113, and carbon tetrachloride (CCl4) were drawn from 2378 of the 4192 bottles collected during the expedition. Specially designed 12-L Bullister bottles were used on the cruise to reduce CFC contamination. These bottles have the same outer diameter as standard 10-L Niskin bottles, but use a modified end-cap design to minimize the contact of the water sample with the end-cap O-rings after closing. The O-rings used in these water sample bottles were vacuum-baked prior to the first station. Stainless steel springs covered with a nylon powder coat were substituted for the internal elastic tubing provided with standard Niskin bottles. When taken, water samples for CFC and carbon tetrachloride analysis were the first samples drawn from the Bullister bottles. Care was taken to coordinate the sampling of CFCs with other samples to minimize the time between the initial opening of each bottle and the completion of sample drawing. In most cases, dissolved oxygen, 3He, and HCFC samples were collected within several minutes of the initial opening of each bottle. To minimize contact with air, the CFC samples were drawn directly through the stopcocks of the Bullister bottles into 100-mL precision glass syringes equipped with two-way metal stopcocks. The syringes were immersed in a holding tank of clean surface seawater until analyzed.

Concentrations of CFC-11 and CFC-12 were measured by shipboard electron capture gas chromatography (EC-GC) using techniques modified from those described by Bullister and Weiss (1988). For seawater analyses, water was transferred from a glass syringe to a fixed volume chamber (~30 mL). The contents of the chamber were then injected into a glass-sparging chamber. The dissolved gases in the seawater sample were extracted by passing a supply of CFC-free purge gas through the sparging chamber for a period of 4 min at 70 mL/min. Water vapor was removed from the purge gas during passage through an 18-cm long, 3/8-in. diameter glass tube packed with the desiccant magnesium perchlorate. The sample gases were concentrated on a cold-trap consisting of a 1/8-in. outside diameter (OD) stainless steel tube with a ~10-cm section packed tightly with Porapak N (60-80 mesh). A vortex cooler, using compressed air at 100 psi, was used to cool the trap to approximately −20°C. After 4 min of purging, the trap was isolated, and the trap was heated electrically to ~100°C. The sample gases held in the trap were then injected onto a precolumn (~25 cm of 1/8-in. OD stainless steel tubing packed with 80100 mesh Porasil C, held at 70°C) for the initial separation of CFC-12, -11, and -113 from carbon tetrachloride. After the CFCs had passed from the pre-column into the main analytical column (~183 cm of 1/8-in.OD stainless steel tubing packed with Carbograph 1AC, 80-100 mesh, held at 70°C) of GC1 (a HP 5890 Series II gas chromatograph with an EC detector), a valve was used to direct the precolumn flow (and more slowly eluting carbon tetrachloride peak) to a second gas chromatograph (Shimadzu Mini II GC with ECD). For the first 52 stations, the chromatographic column in the Shimadzu GC was 1 m of 1/8-in. OD stainless steel tubing packed with 80-100 mesh Porasil C.

Both of the analytical systems were calibrated frequently, with frequency listed below, using a standard gas of known CFC composition. Gas sample loops of known volume were thoroughly flushed with standard gas and injected into the system. The temperature and pressure were recorded so that the amount of gas injected could be calculated. The procedures used to transfer the standard gas to the trap, precolumn, main chromatographic column, and EC detector were similar to those used for analyzing water samples. Two sizes of gas sample loops were used. Multiple injections of these loop volumes could be made to allow the system to be calibrated over a relatively wide range of concentrations. Air samples and system blanks (injections of loops of CFC-free gas) were injected and analyzed in a similar manner. The typical analysis time for seawater, air, standard, or blank samples was ~11 min.

Concentrations of the seawater samples, and gas standards are reported relative to the SIO98 calibration scale. Dissolved CFC concentrations are given in units of picomoles per kilogram seawater (pmol/kg). CFC concentrations in seawater samples were determined by fitting their chromatographic peak areas to multi-point calibration curves, generated by injecting multiple sample loops of gas from a working standard (PMEL cylinder 45191 for CFC-11, CFC-12, CFC-113, and CCl4) into the analytical instrument. The response of the detector to the range of moles of CFC passing through the detector remained relatively constant during the cruise. Full-range calibration curves were run at intervals of 14 days during the cruise. These were supplemented with occasional injections of multiple aliquots of the standard gas at more frequent time intervals. Single injections of a fixed volume of standard gas at one atmosphere were run much more frequently (at intervals of ~90 minutes) to monitor short-term changes in detector sensitivity. The CFC-113 peak was often on a small bump on the baseline, resulting in a large dependence of the peak area on the choice of endpoints for integration. To provide better precision, the height of the peak was used instead . The precisions of measurements of the standard gas in the fixed volume (n = 690) were ± 0.67% for CFC-12 and 0.59% for CFC-11.

The efficiency of the purging process was evaluated periodically by re-stripping high- concentration surface water samples and comparing the residual concentrations to initial values. These re-strip values ranged from approximately 1% for CFC-11 and CFC-12 in cold waters to values of <1% in warm waters. A fit of the re-strip efficiency as a function of temperature was applied to the final data set. The cold-water values have been applied to all values in the preliminary data set.

There were very few measurements of CFC-11 and CFC-12 concentrations less than 0.005 pmol/kgalong A16S section.

Based on the analysis of 100 duplicate samples, the precisions (1 standard deviation) of 0.45% or 0.003 pmol/kg (whichever is greater) for dissolved CFC-11, and 0.78% or 0.004 pmol/kg for CFC-12 measurements was estimated.

A very small number of water samples had anomalously high CFC concentrations relative to adjacent samples. These samples occurred sporadically during the cruise and were not clearly associated with other features in the water column (e.g., anomalous dissolved oxygen, salinity, or temperature features). This suggests that these samples were probably contaminated with CFCs during the sampling or analysis processes. Measured concentrations for these anomalous samples are included in the data set, but are given a quality flag value of either 3 (questionable measurement) or 4 (bad measurement). A quality flag of 5 was assigned to samples drawn from the rosette but never analyzed due to various reasons (e.g., leaking stopcock, plunger jammed in syringe barrel). A total of 13 analyses of CFC-11 and 16 analyses of CFC-12 were assigned a quality flag 3. Four analyses of CFC-11 and 6 analyses of CFC-12 were assigned a quality flag 4. A total of 13 samples were given a quality flag 5 (sampled but not analyzed).

3.8 Underway pCO2 (fCO2) Measurements

The shipboard automated underway pCO2 system was situated in the hydrographic laboratory. It ran on an hourly cycle during which 3 gas standards, 8 headspace samples from the equilibrator, and 3 ambient air samples were analyzed. The system consisted of an equilibrator where surface seawater from the bow intake was equilibrated with headspace, a valve box that contained the infrared analyzer, and an electronics box with a computer and interface boards that controlled valves and log sensors.

The equilibrator, designed by R. Weiss of SIO, is made from a 58 cm H X 23 cm inside diameter (ID) cylindrical PlexiglasTM chamber. Surface seawater flows through a showerhead in the top at a rate of 10-13 L/min. The water spray through the 16-L headspace and the turbulence of the water streams impinging on the surface of 8 L of water cause the gases in water and headspace to equilibrate. Excess water flows through an outlet 20 cm from the bottom of the equilibrator into an over-the-side drain. Two vents in the top of the equilibrator ensure that the headspace remains at the measured laboratory pressure. Headspace gas circulates in a closed loop driven by a KNF pump at 200 mL/min. From the equilibrator, the gas passes through a condenser, a column of magnesium perchlorate, a mass flow meter (MFM), the 12 mL sample cell of a LicorTM Model 6251 non-dispersive infrared analyzer (IR), and back into the equilibrator headspace.

A second KNF pump draws marine air from an intake on the bow mast through ~100 m of 0.95 cm (3/8in.) OD DekoronTM tubing at a rate of 68 L/min. A filter of glass wool at the intake prevents particles from entering the gas stream. At designated times, the program diverts ~250 mL/min of air from this line into the LicorTM sample cell for analysis. Excess marine air and the two vent lines from the equilibrator empty through an endcap into an open-ended PVC tube. This means that any air drawn into the equilibrator is marine air rather than lab air with elevated and variable CO2 concentration.

Both sample streams (equilibrator headspace and marine air) are analyzed bone dry. They pass first through a cold trap (condensor) at 5°C and then through a column of magnesium perchlorate. Standard gases also run through the magnesium perchlorate.

A custom developed program run under LabView controls the system and graphically displays air and water mole-fraction of CO2 (xCO2) readings. The program logs the voltage and temperature of the infrared analyzer, water flow, gas flows, equilibrator temperature, and barometric pressure. It also logs temperature and salinity from a Seabird MicroTSG unit connected to the seawater line near the equilibrator. An RS-422 feed from the shipboard computing system (SCS) provides additional data including time, latitude, longitude, temperature, and salinity in the sea chest near the seawater intake, relative and absolute wind speed and direction, and fluorometer readings. The program writes all of this data to disk at the end of each measurement phase.

The details of instrumental design can be found in Wanninkhof and Thoning (1993), Ho et al. (1995), and Feely et al. (1998).

The system runs on an hourly cycle during which 3 standard gases, 3 marine air samples, and 8 surface water samples (from the equilibrator headspace) are analyzed on the schedule listed in Table 3.3. A Valco multi-port valve selects the gas to be analyzed. Each measurement phase starts by flowing either standard (at ~50mL/min), equilibrator headspace (at ~200 mL/min), or marine air (at ~250 mL/min) through the Licor. Fifteen seconds before the end of each phase, a solenoid valve stops the gas flow. Ten seconds later, the program logs all sensors and writes the data to disk.

Table 3.4. Hourly sampling cycle for the underway pCO2 system (version 2.5).
Minutes after the Hour Sample
4 Low standard
8 Mid standard
12 High standard
16.5 Water (= headspace of equilibrator)
21 Water
25.5 Water
30 Water
34 Air (marine air from the bow line)
38 Air
42 Air
46.5 Water
51 Water
55.5 Water
60 Water

The headspace equilibration time, as determined by return to equilibrium after perturbation by adding nitrogen to the headspace, was approximately 2.5 min. The transit time of water from the bow to the equilibrator was determined in 1998 by injecting a slug of dye into the intake and measuring the response on a fluorometer that is located in the hydrolab, close to the equilibrator. The response time, defined as the time elapsed between peak concentration and the half peak level, t1/2, was 1.45 min. This short time suggests little dispersion of the water during transit through the tubing.

The unit is standardized every hour with three compressed air standards containing known amounts of CO2 gas in (natural) air. The standard gases were purchased from NOAA/CMDL in Boulder and are directly traceable to the World Meteorological Observatory (WMO) scale.

The standards used on the cruise were:

Tank No. Mole Fraction CO2 (ppm) (xCO2)
CC71588 531.98
CA05344 411.42
CA05395 315.25

All xCO2 values are reported in parts per million (ppm), and fCO2 values are reported in micro atmospheres (atm).

The mixing ratios of ambient air and equilibrated headspace air are calculated by fitting a second-order polynomial through the hourly-averaged voltage response of the detector versus mixing ratios of the standards preceding and following the air and water samples. Mixing ratios of dried equilibrated headspace and air are converted to fCO2 in surface seawater and water saturated air. For ambient air (a) and equilibrator headspace (eq), the fCO2a, or fCO2eq are calculated assuming 100% water vapor content:

fCO2a/eq = xCO2a/eq(P−pH2O) exp[(B11+2d12)P/RT]

where fCO2a/eq is the fugacity in ambient air or equilibrator, pH2O is the water vapor pressure at the sea surface temperature, P is the atmospheric pressure (in atm), T is the sea surface temperature (SST) or equilibrator temperature (in K), and R is the ideal gas constant (82.057 mL atm deg-1 mol-1). The exponential term is the fugacity correction where B11 is the second viral coefficient of pure CO2:

B11 = −1636.75 + 12.0408T − 0.032795T2 + 3.16528 X 10-5 T3

and

d12 = 57.7 − 0.118 T

where d12 is the correction for an air-CO2 mixture in units of mL/mol (Weiss 1974).

The calculation for the fugacity at SST as measured by the thermosalinograph involves a temperature correction term for the increase of fCO2 due to heating of the water from its passage through the pump and through 5-cm ID Teflon-sleeved stainless steel tubing within the ship. The water in the equilibrator is typically 0.2 to 0.3°C warmer than sea surface temperature. At the southern end of the transect when SST ≈ 25°C, the difference was as much as 1°C. The empirical temperature correction from equilibrator temperature to SST is outlined in Weiss et al. (1982).

dln(fCO2) = (teq − SST) X [0.03107 − 2.7851 X 10-4 teq − 1.839 X 10-3 ln(fCO2eq)]

where dln(fCO2) is the difference between the natural logarithm of the fugacity at teq and SST, and teq is the equilibrator temperature in °C.

The precision of the measurements is estimated at 0.2 ppm based on repetitive measurements of marine air. The accuracy of the air values is believed to be better than 0.5 ppm based on comparisons with flask samples on tests performed in 1995. Equilibrator headspace values are believed to be accurate to within 2 atm. The greater uncertainty is attributed to the equilibration efficiency. Outside the calibration range of the standards, an accuracy of 5 ppm (atm) is assigned based on laboratory tests where the calibrated IR output is compared with standards of known concentration outside the calibration range.

At the start of the cruise, from 1/14/05 17:04 (UTC) until 1/16/05 15:50 (UTC), problems with the electrical power caused the program to run out of sync and the equilibrator thermistor to give bad readings. Connecting the instrument to an uninterruptible power supply (UPS) eliminated the problem.

4. How to Obtain the Data and Documentation

This database (NDP-087) is available free of charge from CDIAC. The data are available from CDIAC's anonymous file transfer protocol area. The complete documentation and data can be obtained from the CDIAC oceanographic Web site.

For additional information, contact CDIAC.

5. References

  • Armstrong, F.A.J., C.R. Stearns, and J.D.H. Strickland. 1967. The measurement of upwelling and subsequent biological processes by means of the Technicon AutoAnalyzer and associated equipment. Deep-Sea Research 14:381-89
  • Bradshaw, A. L., and P. G. Brewer. 1988. High-precision measurements of alkalinity and total carbon dioxide in seawater by potentiometric titration-1: Presence of unknown protolyte(s)? Maine Chemistry 23:69-86.
  • Bullister, J.L., and R.F. Weiss. 1988. Determination of CC13F and CC12F2 seawater and air. Deep-Sea Research 25:839-53.
  • Byrne, R. H., and J. A. Breland. 1989. High-precision multi-wavelength pH determinations in seawater using cresol red. Deep-Sea Research 36:803-10.
  • Carpenter, J. H. 1965. The Chesapeake Bay Institute technique for the Winkler dissolved oxygen method. Limnology and Oceanography 10:141-43.
  • Chipman, D. W., J. Marra, and T. Takahashi. 1993. Primary production at 47N and 20W in the North Atlantic Ocean: A comparison between the 14C incubation method and mixed layer carbon budget observations. Deep-Sea Research II 40:151-69.
  • Clayton T., and R. H. Byrne. 1993. Calibration of m-cresol purple on the total hydrogen ion concentration scale and its application to CO2-system characteristics in seawater. Deep-Sea Research 40:211-529.
  • Culberson, C. H., G. Knapp, M. Stalcup, R. T. Williams, and F. Zemlyak. 1991. A Comparison of Methods for the Determination of Dissolved Oxygen in Seawater. WHP Office Report, WHPO 91-2. WOCE Hydrographic Program Office, Woods Hole, Mass.
  • Dickson, A. G. 1981. An exact definition of total alkalinity and a procedure for the estimation of alkalinity and total CO2 from titration data. Deep-Sea Research 28:609-23.
  • Dickson, A.G. 1990. Thermodynamics of the dissociation of boric acid in synthetic seawater from 273.15 to 318.15 K. Deep-Sea Research Part A, 37/5:755-66.
  • Dickson, A.G. 1993. The measurement of seawater. Marine Chemistry 44/24:131-42.
  • Dickson, A.G., and F.J. Millero. 1987. A comparison of the equilibrium constants for the dissociation of carbonic acid in seawater media. Deep-Sea Research Part A, 34/10:173-343.
  • Dickson, A.G., and J.P. Riley. 1979. The estimation of acid dissociation constants in seawater media from potentiometric titration with strong base, 1: The ionic product of water-KSUS-w. Marine Chemistry 7/2:89-99.
  • DOE (U.S. Department of Energy). 1994. Handbook of Methods for the Analysis of the Various Parameters of the Carbon Dioxide System in Seawater. Version 2.0. ORNL/CDIAC-74. Ed. A. G. Dickson and C. Goyet. Carbon Dioxide Information Analysis Center, Oak Ridge National Laboratory, Oak Ridge, Tenn.
  • Forde, E. B., J.C. Hendee, and R. Wanninkhof. 1994. Hydrographic, carbon dioxide, nutrient, and productivity measurements from the South Atlantic during July and August of 1991, NOAA Data Report ERL AOML-24: NOAA/AOML
  • Friederich, G. E., P. Sherman, and L. A. Codispoti. 1984. A High-Precision Automated Winkler Titration System Based on a HP-85 Computer, A Simple Colorimeter and an Inexpensive Electromechanical Buret. Bigelow Laboratory Technical Report, 42. Bigelow Laboratory for Ocean Sciences, West Boothbay Harbor, Maine.
  • Gordon, L.I., J.C. Jennings Jr., A.A. Ross, and J.M. Krest. 1993. A suggested protocol for continuous automated analysis of seawater nutrients (phosphate, nitrate, nitrite, and silicic acid) in the WOCE Hydrographic program and the Joint Global Ocean Fluxes Study. WOCE Operations Manual, Vol. 3: The Observational Programme, Section 3.2: WOCE Hydrographic Programme, Part 3.1.3: WHP Operations and Methods. WHP Office Report WHPO 91-1; WOCE Report No. 68/91. November, 1994, Revision 1, Woods Hole, Mass., USA, 52 pp.
  • Ho, D.T., R. Wanninkhof, J. Masters, R.A. Feely, and C.E. Cosca. 1997. Measurement of underway fCO2 in the eastern equatorial Pacific on NOAA ships Baldrige and Discoverer. NOAA Data Report, ERL AOML-30, 52 pp.
  • Johnson, K. M. 1992. Operators Manual: Single-Operator Multiparameter Metabolic Analyzer (SOMMA) for Total Carbon Dioxide (CT) with Coulometric Detection. Brookhaven National Laboratory, Brookhaven, N.Y.
  • Johansson, O., and M. Wedborg. 1982. On the evaluation of potentiometric titrations of seawater with hydrochloric acid. Oceanologica Acta 5:209-18.
  • Johnson, K. M., A. E. King, and J. McN. Sieburth. 1985. Coulometric TCO2 analyses for marine studies: An introduction. Marine Chemistry 16:61-82.
  • Johnson, K. M., A. Kotzinger, L. Mintrop, J. C. Duinker, and D. W. R. Wallace. 1999. Coulometric total carbon dioxide analysis for marine studies: Measurement and internal consistency of underway surface TCO2 concentrations. Marine Chemistry 67:123-44.
  • Johnson, K. M., P. J. Williams, L. Brandstrom, and J. McN. Sieburth. 1987. Coulometric total carbon analysis for marine studies: Automation and calibration. Marine Chemistry 21:117-33.
  • Johnson, K. M., K. D. Wills, D. B. Butler, W. K. Johnson, and C. S. Wong. 1993. Coulometric total carbon dioxide analysis for marine studies: Maximizing the performance of an automated gas extraction system and coulometric detector. Marine Chemistry 44:167-87.
  • Lewis, E., and D.W.R. Wallace. 1998. Program developed for CO2 system calculations. ORNL/CDIAC-105. Carbon Dioxide Information Analysis Center, Oak Ridge National Laboratory, U.S. Department of Energy, Oak Ridge, Tennessee.
  • Marinenko, G., and J.K. Taylor. 1968. Electrochemical equivalents of benzoic and oxalic acid. Analytical Chemistry 40:1645-51.
  • Mehrbach, C., C.H. Culberson, J.E. Hawley, and R.M. Pytkowicz. 1973. Measurement of the apparent dissociation constants of carbonic acid in seawater at atmospheric pressure. Limnology and Oceanography 18:897-907
  • Millero, F. J., A. G. Dickson, G. Eischeid, C. Goyet, P. R. Guenther, K. M. Johnson, K. Lee, E. Lewis, D. Purkerson, C. L. Sabine, R. Key, R. G. Schottle, D. R. W. Wallace, and C. D. Winn. 1998. Assessment of the quality of the shipboard measurements of total alkalinity on the WOCE Hydrographic Program Indian Ocean CO2 survey cruises, 19941996. Marine Chemistry 63:9-20.
  • Millero F. J., J. Z. Zhang, K. Lee, and D. M. Campbell. 1993a. Titration alkalinity of seawater. Marine Chemistry 44:153-65.
  • Millero, F.J., R.H. Byrne, R. Wanninkhof, R. Feely, T. Clayton, P. Murphy, and M.F. Lamb. 1993b. The internal consistency of CO2 measurements in the equatorial Pacific. Marine Chemistry 44:269-80.
  • Peng, T.-H., T. Takahashi, W. S. Broecker, and J. Olafsson. 1987. Seasonal variability of carbon dioxide, nutrients and oxygen in the northern North Atlantic surface water: Observations and a model. Tellus 39B:439-58.
  • Ramette, R.W., C.H. Culberson, and R.G. Bates. 1977. Acid-base properties of Tris(hydroxymethyl)aminomethane (Tris) buffers in seawater from 5 to 40°C. Analytical Chemistry 49:867-70.
  • Taylor, J.K., and S.W. Smith. 1959. Precise coulometric titration of acids and bases. Journal Research of National Bureau of Standards 63:153-59.
  • UNESCO, 1981. Background papers and supporting data on the practical salinity scale 1978. UNESCO Technical Papers in Marine Science, No. 37, p. 144.
  • Wanninkhof, R., and K. Thoning. 1993. Measurement of fugacity of CO2 in surface water using continuous and discrete sampling methods. Marine Chemistry 44:189-205.
  • Wanninkhof, R. and S.C. Doney, 2005: Cruise Report CLIVAR A16S_2005, http://whpo.ucsd.edu/data/co2clivar/atlantic/a16/a16s_2005a/index.htm
  • Weiss, R.F. 1974. Carbon dioxide in water and seawater: The solubility of a non-ideal gas. Marine Chemistry 2:203-15.
  • Weiss, R.F., R.A. Jahnke, and C.D. Keeling. 1982. Seasonal effects of temperature and salinity on the partial pressure of CO2 in seawater. Nature 300:511-13.