Information on CTD raw data observed by R/V MIRAI
(Cruise number : MR00-K03)
1. Preface
----------
CTD(Conductivity-Temperature-Depth profiler) is used to observe
the vertical profiles of temperature and conductivity.
Usually, this system is operated with multicylinder water sampler.
Observed signal is transmitted from sensor to the operation room on
board using wire cable, and electric power is supplied from vessel
to sensor.
Details of sensors attached to CTD system for MR00-K03 cruise are
presented in section 3.
The following software, developed and supplied by the Sea-Bird
Electronics, Inc., was used in MR00-K03.
SEASAVE(ver 4.232) for data acquisition
SEASOFT(ver 4.232) for data processing
Data presented on this website is averaged over 1db.
2. CTD sampling system
----------------------
Followings are the options for CTD/RMS(Rosette Multi Sampler) systems;
system1 : CTD/RMS and 24 Niskin bottles of 30L
system2 : CTD/RMS and 12 Niskin bottles of 12L
3. Sensors
----------
CTD/RMS system1
(1) Temperature sensor
Model : SBE3-04F, Sea-Bird Electronics,Inc.
Serial number : 031359
Measurement range : -5.0 to +35degC
Accuracy : 0.001degC
Resolution : 0.0002degC
(2) Salinity sensor
Model : SBE4-04/O, Sea-Bird Electronics,Inc.
Serial number : 041206
Measurement range : 0.0 to 7S/m
Accuracy : 0.0003S/m
Resolution : 0.00004S/m
(3) Pressure sensor
Model : SBE9plus, Sea-Bird Electronics,Inc.
Serial number : 42423
Measurement range : up to 10500m
Accuracy : 0.015%F.S.
Resolution : 0.001%F.S.
(4) Fluorometer
Model : SEA TECH,Inc.
Serial number : 132s
(5) Altimeter
Model : Benthos 2110-1,
Serial number : 0206
CTD/RMS system2
(1) Temperature sensor
Model : SBE3-04F, Sea-Bird Electronics,Inc.
Serial number : 031525
Measurement range : -5.0 to +35degC
Accuracy : 0.001degC
Resolution : 0.0002degC
(2) Salinity sensor
Model : SBE4-04/O, Sea-Bird Electronics,Inc.
Serial number : 041088
Measurement range : 0.0 to 7S/m
Accuracy : 0.0003S/m
Resolution : 0.00004S/m
(3) Pressure sensor
Model : SBE9plus, Sea-Bird Electronics,Inc.
Serial number : 751190
Measurement range : up to 10500m
Accuracy : 0.015%F.S.
Resolution : 0.001%F.S.
(4) Altimeter
Model : Datasonics PSA-900D,
Serial number : 396
4. Notes
--------
(1) Observed data by system1 is stored in the data file, its filename
includes the letter of 'L' or 'l'.
Observed data by system2 is stored in the data file, its filename
includes the letter of 'S' or 's'.
(2) Data processing sequence for SEASOFT as follows;
--------------------------------------------------------------------
command function
-------- ----------------------------------------------------------
datcnv Convert raw data to engineering units, and store converted
data in file.
section Extract rows of data from file.
wildedit Mark a data value with badflag to eliminate wild points.
binavg Average data, basing bins on pressure, depth, scan number,
or time range.
split Split data in file into upcast and downcast files.
--------------------------------------------------------------------