1.0 Scope and field of application
This procedure describes a method for
the determination of total dissolved inorganic carbon in sea water, expressed
as moles of carbon per kilogram of sea water. The method is suitable for
the assay of oceanic levels of total dissolved inorganic carbon (1800–2300
mmol·kg
–1 ) and also for higher levels such as are found in the Black
Sea (3800–4300 mmol·kg
–1 ). For a more definitive and comprehensive treatment of the analysis,
the reader is referred to the D.O.E. (1991) handbook (Dickson, A.G., and
Goyet, C., eds.) and the
SOMMA manual (Johnson, 1992). The D.O.E.
(1991) handbook by Dickson and Goyet provides protocols for other carbon
dioxide system parameters (i.e. pH, TA, pCO2 ).Scientists who
employ this or other methods to measure total inorganic carbon should make
themselves aware of the current and historical issues that surround these
techniques and make appropriate decisions about specific methodologies
for their application based on the scientific requirements and constraints
of their individual programs.
2.0 Definition
The total dissolved inorganic carbon content
of seawater is defined as:
3.0 Principle of Analysis
Total dissolved inorganic carbon (C T
) is measured by acidifying a seawater sample to convert HCO3 -
and CO32- to undissociated CO2 , and then
extracting this CO2 as a gas, trapping and titrating the amount
evolved (Johnson et al., 1987). A high degree of precision and accuracy
is maintained by using a computer-controlled automated dynamic headspace
analyzer that extracts total carbon dioxide (CT ) from seawater
using a SOMMA (Single-Operator Multiparameter Metabolic Analyzer) designed
by K. Johnson of Brookhaven National
Laboratory (Johnson, 1992). This apparatus is coupled to a commercial coulometer
that detects the extracted CO2 (Huffman, 1977; Lindberg and
Cedergren, 1978).
The analytical system forces a sample (either seawater, Na2CO3 standard solution or distilled water) into a volume-calibrated pipette using a pressurized headspace gas of pure N2 . This known sample volume (~30 ml) is then dispensed into a stripping chamber previously acidified with ~1.5 ml of phosphoric acid (the chamber and phosphoric acid are purged with pure N 2 carrier gas prior to the addition of each sample). CO2 gas evolved from the acidified sample is then passed through a thermostated condenser (4°C) and a magnesium perchlorate trap to remove water vapor. Any acidic or reactive gases are removed by passing through activated silica gel.
The amount of CO2 gas extracted
from the acidified sample by a continuous flow of pure N2 through
the chamber is determined coulometrically by trapping and titrating the
CO2 with a DMSO based absorbent containing ethanolamine. The
resulting hydroxyethylcarbamic acid which is formed with electrochemically
generated hydroxide ions is titrated to maintain the absorbing solution
at constant pH. Relevant chemical
equations occurring in the solution are:
The determination of seawater total dissolved inorganic carbon is calibrated with known volumes of pure CO 2 (Johnson, 1992), a modification of the CO 2 gas loop system described by Johnson et al. (1985) and Johnson et al. (1987).
4.0 Apparatus
4.1 SOMMA (Single-Operator Multi-Parameter Metabolic Analyzer), available from University of Rhode Island in conjunction with K. Johnson, Brookhaven National Laboratory. Includes:
4.1.1 Eight port gas chromatography valve
4.1.2 Two loops of stainless steel tubing of known volume
4.1.3 Three thermistors accurate to ± 0.05°C
4.1.4 SOMMA glassware (e.g. stripping chamber, calibrated water-jacketed pipette, water-cooled condenser, aerosol traps)
4.1.5 Temperature controlled water bath circulators
4.1.6 Temperature controlled sample bottle holder
4.2 A model 5011 CO 2 coulometer (UIC Inc., P.O. Box 863, Joliet, IL 60434)
4.2.1 Coulometer cell (temperature controlled)
4.2.2 Electrodes: platinum spiral cathode and silver rod anode
4.2.3 Rubber cell top
4.2.4 Stir bar
4.3 Computer system
4.3.1 PC (e.g. 286)
4.3.2 Printer
4.3.3 Software program (K. Johnson)
4.4 Sampling equipment
4.4.1 Clean 1000 ml borosilicate reagent bottles and ground-glass stoppers
4.4.2 Apiezon L grease
4.4.3 Tygon Ò drawing tube
4.4.4 Pipette to dispense mercuric chloride
4.5 Other: Barometer, e.g. Paroscientific transducer
5.0 Reagents
5.1 Compressed gases:
5.1.1 Carrier gas: Nitrogen (>99.9995%)
5.1.2 Calibration gas: CO 2 (>99.999%)
5.1.3 Headspace gas: Air (350 matm CO2 )
5.2 Phosphoric acid (reagent grade): Phosphoric acid, diluted with deionized water by a factor of 10:1 (~8%) is used to acidify seawater samples.
5.3 Magnesium perchlorate (reagent grade): For the removal of water vapor.
5.4 Activated silica gel: For the removal of reactive acidic gases. Glass tubes (ORBO-53 traps) with activated silica are custom-made by Supelco Inc., U.S.A.
5.5 Cathode solution: UIC Coulometrics, Inc. proprietary mixture of water, ethanola-mine, tetraethylammonium bromide, and thymolphthalein in solution in dimethyl sulfoxide (DMSO).
5.6 Anode solution: UIC Coulometrics, Inc. proprietary solution containing saturated potassium iodide in water and DMSO.
5.7 Potassium iodide (reagent grade): Added to anode solution.
5.8 Saturated solution of Mercuric chloride:
5.9 Ascarite: For the removal of CO 2 from the carrier gas.
5.10 Sodium carbonate (optional): Na 2 CO 3 (99.95% pure: Alkimetric standard, Fisher Scientific Co.): Six solutions are prepared for standard calibration ranging in con-centration from distilled water to 500, 1000, 1500, 2000, and 2500 mmol C.kg -1 .
6.0 Sampling
6.1 Seawater sample for C T analysis are
collected in the teflon-coated Niskin bottles either on the General Oceanics
rosette or individually mounted on stainless steel hydrowire. C T seawater
is collected after the first and replicate oxygen samples.
6.2 The samples are drawn into 12 individually
numbered, clean, one liter borosilicate glass bottles. Water is also drawn
into at least three duplicate bottles. In obtaining seawater samples, care
is taken to minimize turbulence and to prevent the retention of air bubbles
in the bottles. A 30-50 cm length of Tygon Ò tubing is connected
to the Niskin bottle spout. The end of the tube is elevated before the
spout is opened to pre-vent
the trapping of bubbles in the tube. With
the water flowing, the tube is placed in the bottom of the bottle. The
bottle is slowly rotated and the side of the bottle tapped with the stopper
to ensure that no air bubbles adhere to the bottle walls. At least two
to three volumes of water are allowed to overflow from the bottle. A headspace
of >1% of the bottle volume is left to allow for water expansion. 200 ml
of saturated mercuric chloride is then added to the sample bottle to prevent
further biological activity. The bottle neck is dried with a Kim-Wipe stick
and then the bottle is sealed with an Apiezon grease ground-glass stopper,
ensuring that it remains gas-tight.
Rubber bands are placed around the lip
of bottle and the stopper in crisscross manner as positive closure of the
bottle.
6.3 The samples are then stored in a cool, dark location until analysis.
7.0 Procedures
7.1 Bottle preparation: Bottles should
be carefully cleaned before use. Used bottles are emptied and any grease
on the bottle neck is removed with kim-wipes. Bottles are thoroughly washed
with a commercial detergent, then rinsed with a 10% HCl solu-tion. Copious
rinsing with deionized/distilled water is followed by an acetone rinse.
The bottles are then allowed to air dry for an hour and sealed with ground-glass
stop-pers.
7.2 Maintenance of SOMMA-Coulometer system
7.2.1 The titration cell is cleaned with copious rinses of deionized water and a rinse with acetone. The idearm of the cell is then filled with acetone which is then left to drain overnight through the frit separating the cathode compartment from the sidearm. The cell is then rinsed with deionized water and left to dry overnight at 55°C.
7.2.2 The rubber top, electrodes, stir bar and perchlorate trap are thoroughly cleaned with deionized water.
7.2.3 The titration cell, rubber stopper, stir bar, electrodes, magnesium perchlorate glass trap, and teflon carrier gas lines are dried overnight at 55°C.
7.2.4 The ORBO-53 tubes and magnesium perchlorate traps are renewed with each newly prepared coulometer cell.
7.3 Determination of the background level
7.3.1 Each analysis session, fresh coulometer cell solutions are used.
7.3.2 An aliquot of phosphoric acid is
introduced into the coulometer cell. CO 2 -free N 2 carrier gas is allowed
to run through the SOMMA system and into
coulometer cell. Once the background titration rate is stable, a background
level is determined by averaging over a 10 minute period.
7.4 Calibration
7.4.1 The electrical calibration of the
coulometer is not perfectly accurate and the current efficiency of the
electrode processes occurring in the coulometer
cell have been shown to vary from 100% (D.O.E., 1991). It is therefore
necessary to calibrate the coulometer using known volumes of pure CO 2
or with a suite
of Na 2 CO 3 solutions (e.g. Goyet and
Hacker, 1992). The amount of C T titrated by the coulometer is recorded
by microcomputer.
7.4.2 Valves that operate the CO 2 calibration and sample extraction systems are controlled by microcomputer (Johnson 1992).
7.4.3 A stainless-steel loop of known volume is filled with pure CO 2 . The loop is then flushed with carrier gas into the coulometer cell and titrated. A mean calibration factor is calculated from two different loops.
7.4.4 Standard concentrations of C T ranging
from 500 to 2500 mmol.C kg -1 can also be prepared using distilled water
and variable amounts of dried Na 2 CO 3 salt (D.O.E., 1991; Goyet and Hacker,
1992). Sodium carbonate solutions are treated as if they were seawater
samples. A blank standard solution (distilled
water without any Na 2 CO 3 ) is also
prepared.
7.5 Analysis of a seawater sample
7.5.1 Once the background level and calibration factor have been determined satis-factorily, the coulometric system can be used to analyze seawater samples.
7.5.2 The stripping chamber is drained on any previous sample. An aliquot of phosphoric acid (~1.5 ml) is dispensed into the stripping chamber. The pipette and the silicone sample lines are flushed with a new sample. The pipette is then filled and allowed to drain into the stripping chamber. CO 2 gas evolved is transferred with carrier N 2 gas to the coulometer cell and titrated.
7.6 Post-analysis: The dispensing and stripping systems are cleaned by rinsing with deionized/distilled water.
8.0 Calculation and expression of results
The amount of C T titrated by the coulometer
for a seawater sample is multiplied by the calibration factor (slope of
calculated vs. measured concentration), giving the seawater C T , expressed
in mmol kg -1 of seawater is computed as follows:
Where:
C = total dissolved inorganic carbon (mmol.kg
-1 )
N = coulometer reading in counts
b = background level of the system
c = coulometer calibration factor
t = time required to measure pipette
V = volume of seawater sample
p = density of seawater
9.0 Quality assurance
9.1 Quality control: Written instructions
outlining the standard operating procedures are maintained and continually
reviewed and updated. Standard operating procedures are kept within guidelines
proposed by the D.O.E. CO 2 survey science team (D.O.E., 1991).
9.2 Quality assessment:
9.2.1 The background level is usually within 0.1 mg C min -1 .
9.2.2 The recovery of CO 2 compared to theory during gas calibration is maintained at better than 0.2%. The two gas loops must give the same calibration factor within 0.05% before analysis of samples begin.
9.2.3 Within-bottle replicate and between-bottle duplicate measurement of over 100 samples give a standard deviation of approximately 0.3 mmol.kg -1 and 0.5 mmol.kg -1 respectively, well within guidelines proposed (D.O.E., 1991).
9.2.4 Stable seawater reference materials, supplied by A. Dickson, are analyzed regularly. Analyses of these reference materials at BBSR are within the standard deviation of the mean reported by the Scripps Institution of Oceanography (0.3 mmol.kg -1 ). Intercomparrison exercises are also undertaken with other laboratories.
10.0 References
D.O.E. (1991). Handbook of methods for
the analysis of the various parameters of the carbon dioxide system in
seawater; version 1.0, edited by A.G. Dickson and C. Goyet.
Goyet, C., and Hacker, S.D. (1992). Procedure
for calibration of a coulometric system used for total inorganic carbon
measurements in seawater, Marine Chemistry, 38, 37-51.
Huffman, E.W.D. Jr. (1977). Performance
of a new automatic carbon dioxide coulometer. Microchemical Journal, 22,
567-573.
Johnson, K.M. (1992). Single-Operator
Multiparameter Metabolic Analyzer (SOMMA) for Total Carbon Dioxide (C T
) with Coulometric Detection. SOMMA Manual 1.0, January 1992, Brookhaven
National Laboratory, 70 pp.
Johnson, K.M., King, A.E., and Sieburth,
J.McN. (1985). Coulometric TCO 2 analyses for marine studies: an introduction.
Marine Chemistry, 16, 61-82.
Johnson, K.M., Sieburth, J.M., Williams,
P.J. leB., and Brandstrom, L. (1987). Coulometric total carbon dioxide
analysis for marine studies: automation and calibration. Marine Chemistry,
21: 117-133.
Lindberg, A.O., and Cedergren, A. (1978).
Automatic coulometric titration with photometric endpoint detection. Part
II. Coulometric determination of nanomolar amounts of carbon dioxide by
non-aqueous titration. Analytica Chimica Acta, 96, 327- 333.
UNESCO. (1991). Reference materials for
oceanic carbon dioxide measurements. In: Report of the Sub-panel on Standards
for CO 2 Measurements of the Joint Panel on Oceanographic Tables and Standards.
Report of meeting in Acapulco, Mexico, 29-31 August 1988. UNESCO Technical
Papers in Marine Science.